The standard sample of the propanol-1 solution composition is intended for:
- Certification and error control of measurement methods for propanol-1 content in biological objects, environmental objects and technical products;
- For verification, calibration and calibration of chromatographs and other, including specialized, measuring instruments (SI) designed to determine propanol-1 in biological objects, environmental objects and technical products, as well as for monitoring the metrological characteristics of SI during their tests, including for type approval purposes.
Features:
CO is an aqueous solution of propanol-1, the range of permissible certified values is from 3.5 g/dm3 to 4.5 g/dm3, packaged in glass ampoules with a volume of at least 3 cm3.
Technical specifications:
- The mass concentration of propanol-1 is 4.00 g/dm3;
- The limits of the relative error of the certified value of CO - ?, at P = 0.95, are ± 1.0 %;
- The standard sample is an aqueous solution of propanol (propane-1-ol), with a mass fraction of the main component (A) of at least 99.7%. .
The shelf life of a copy of CO is 1 year.
Transport data/dimensions/weight:
Glass ampoules of 3 cm3, 5 ampoules are placed in a cardboard box.
Basic equipment:
- A glass ampoule containing 3 ml of GSO solution - 1 pc.;
- Passport- 1 pc.