Purpose: The EСROS-5700 concentrator is designed to measure the content of petroleum products, fats and nonionic surfactants (NPAV) in an extract from samples of waters for various purposes, bottom sediments in laboratories of industrial enterprises, research institutions, control bodies, educational institutions, environmental and ecological analytical laboratories, etc.5700 is registered in the State Register of Measuring Instruments under № 77795-20.
Principle of operation: absorption of infrared radiation by molecules of petroleum hydrocarbons, fats and NPA at a wavelength of 3.42 microns (2930 cm-1) in extracts. Carbon tetrachloride (CCl4) or dichloroethane (C2Cl6) or other halogenated hydrocarbons are used as an extractant.
Metrological and technical characteristics:
The range of indications for the mass concentration of petroleum products, fats, and NSAIDs in extracts: from 0 to 150 mg/dm3
The measurement range of the mass concentration of petroleum products, fats and NSAIDs in extracts: from 0 to 100 mg/dm3
The limits of the permissible absolute error of the concentration meter when measuring the mass concentration of petroleum products or fats: ±(0.5 + 0.04·K) mg/dm3, where K is the mass concentration
of petroleum products or fats, mg/dm3
The limits of the permissible absolute error of the concentration meter when measuring the mass concentration of NPA: ±(1.0 + 0.04·K) mg/dm3, where K is the mass concentration of NPA, mg/dm3
Overall dimensions (L?w?H), mm: 180?180?60
Weight: 1 kg
Power consumption: no more than 9 watts
Power Supply: 207-253 V/49-51 Hz
Average service life: 10 years
The scope of delivery:
EСROS-5700 concentrator -1 piece
Measuring cuvette -1 piece
Network cable - 1 pc
Tripod for chromatographic columns - 1 pc
Chromatographic columns - 6 pcs
Standard sample of the composition of a solution of petroleum products in carbon tetrachloride, GSO 7248-96 – 1 pack (5 ampoules)
User manual, passport - 1 piece
The method of verification of MP 242-2329-2019 - 1 piece
ZIP set – 1 pc